https://doi.org/10.1140/epjti/s40485-015-0023-9
Research article
Novel experimental setup for time-of-flight mass spectrometry ion detection in collisions of anionic species with neutral gas-phase molecular targets
1
Instituto de Física Fundamental, Consejo Superior de Investigaciones Científicas, Serrano 113-bis, 28006, Madrid, Spain
2
Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Avenida Complutense 22, 28040, Madrid, Spain
3
Laboratório de Colisões Atómicas e Moleculares, CEFITEC, Departamento de Física, Faculdade de Ciências e Tecnologia, Universidade Nova de Lisboa, 2829-516, Caparica, Portugal
4
Centre of Medical Radiation Physics, University of Wollongong, Wollongong, NSW, 2522, Australia
* e-mail: g.garcia@csic.es
Received:
15
June
2015
Accepted:
4
August
2015
Published online:
21
August
2015
We report a novel experimental setup for studying collision induced products resulting from the interaction of anionic beams with a neutral gas-phase molecular target. The precursor projectile was admitted into vacuum through a commercial pulsed valve, with the anionic beam produced in a hollow cathode discharge-induced plasma, and guided to the interaction region by a set of deflecting plates where it was made to interact with the target beam. Depending on the collision energy regime, negative and positive species can be formed in the collision region and ions were time-of-flight (TOF) mass-analysed. Here, we present data on O2 precursor projectile, where we show clear evidence of O– and O2 – formation from the hollow cathode source as well as preliminary results on the interaction of these anions with nitromethane, CH3NO2. The negative ions formed in such collisions were analysed using time-of-flight mass spectrometry. The five most dominant product anions were assigned to H–, O–, NO–, CNO– and CH3NO2 –.
Key words: Hollow cathode discharge / Anion beam / Collisions / TOF / Negative ions / Atomic and molecular collisions / Electron transfer
© The Author(s), 2015