Calibration of a multi-anvil high-pressure apparatus to simulate planetary interior conditions
1
Faculty of Science, Vrije Universiteit Amsterdam, Amsterdam, the Netherlands
2
Department of Earth Sciences, Indian Institute of Technology Roorkee, Roorkee, India
* e-mail: j.s.knibbe@vu.nl
Received:
16
April
2018
Accepted:
21
June
2018
Published online:
4
July
2018
This paper presents the setup and pressure calibration of an 800-ton multi-anvil apparatus installed at the Vrije Universiteit (Amsterdam, the Netherlands) to simulate pressure-temperature conditions in planetary interiors. This high-pressure device can expose cubic millimeter sized samples to near-hydrostatic pressures up to ~ 10 GPa and temperatures exceeding 2100 °C. The apparatus is part of the Distributed Planetary Simulation Facility (DPSF) of the EU Europlanet 2020 Research Infrastructure, and significantly extends the pressure-temperature range that is available through international access to this facility.
Key words: High pressure / Multi-anvil / Experimental petrology / Planetary interiors